    Cornea. 2009 Jul;28(6):677-80.

    Evaluation of a new tear osmometer for repeatability and accuracy, using 0.5-microL (500-Nanoliter) samples.

    Yildiz EH, Fan VC, Banday H, Ramanathan LV, Bitra RK, Garry E, Asbell PA.

    Department of Ophthalmology, Mount Sinai School of Medicine, New York, NY 10029, USA.

    PURPOSE: To evaluate the repeatability and accuracy of a new tear osmometer that measures the osmolality of 0.5-microL (500-nanoliter) samples. METHODS: Four standardized solutions were tested with 0.5-microL (500-nanoliter) samples for repeatability of measurements and comparability to standardized technique. Two known standard salt solutions (290 mOsm/kg H2O, 304 mOsm/kg H2O), a normal artificial tear matrix sample (306 mOsm/kg H2O), and an abnormal artificial tear matrix sample (336 mOsm/kg H2O) were repeatedly tested (n = 20 each) for osmolality with use of the Advanced Instruments Model 3100 Tear Osmometer (0.5-microL [500-nanoliter] sample size) and the FDA-approved Advanced Instruments Model 3D2 Clinical Osmometer (250-microL sample size). RESULTS: Four standard solutions were used, with osmolality values of 290, 304, 306, and 336 mOsm/kg H2O. The respective precision data, including the mean and standard deviation, were: 291.8 +/- 4.4, 305.6 +/- 2.4, 305.1 +/- 2.3, and 336.4 +/- 2.2 mOsm/kg H2O. The percent recoveries for the 290 mOsm/kg H2O standard solution, the 304 mOsm/kg H2O reference solution, the normal value-assigned 306 mOsm/kg H2O sample, and the abnormal value-assigned 336 mOsm/kg H2O sample were 100.3, 100.2, 99.8, and 100.3 mOsm/kg H2O, respectively. CONCLUSIONS: The repeatability data are in accordance with data obtained on clinical osmometers with use of larger sample sizes. All 4 samples tested on the tear osmometer have osmolality values that correlate well to the clinical instrument method. The tear osmometer is a suitable instrument for testing the osmolality of microliter-sized samples, such as tears, and therefore may be useful in diagnosing, monitoring, and classifying tear abnormalities such as the severity of dry eye disease.

    PMID: 19512902 [PubMed - indexed for MEDLINE]

    Supplemental Content

    Click here to read

    Patient drug information

    • How to Use Eyedrops

      Wash your hands thoroughly with soap and water.Check the dropper tip to make sure that it is not chipped or cracked.Avoid touching the dropper tip against your eye or anything else—eyedrops and dropper must be kept clean...