    J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Jul 1;877(20-21):1827-32. Epub 2009 May 13.

    Development of a high performance anion exchange chromatography analysis for mapping of oligosaccharides.

    Grey C, Edebrink P, Krook M, Jacobsson SP.

    BioPR&D, AstraZeneca R&D, SE-15185 Södertälje, Sweden. carl.grey@biotek.lu.se

    In the present study a HPAEC-PAD method is described that was developed for monitoring the consistency of N-glycosylation during the production and purification of recombinant proteins and monoclonal antibodies. The method successfully separated 18 neutral and sialylated oligosaccharides. Results obtained were compared with MALDI-TOF MS and it was shown that both methods gave similar results. In addition, a method validation was performed showing that the HPAEC-PAD analysis was well suited for the mapping and characterization of oligosaccharides. The method was found to be robust and additionally the precision was significantly better compared to the MALDI-TOF MS method.

    PMID: 19482525 [PubMed - indexed for MEDLINE]