Enantioselective determination of thyroxine enantiomers by ligand-exchange CE with UV absorbance and ICP-MS detection

Electrophoresis. 2009 May;30(10):1774-82. doi: 10.1002/elps.200800731.

Abstract

A simple CE method has been developed for the separation and determination of thyroxine (T4) enantiomers in pharmaceutical formulations. The method was based on ligand-exchange mechanism using a Cu(II)/L-proline complex as chiral selector. The effects of different parameters affecting separation such as chiral selector concentration, organic additive, buffer pH and temperature were investigated. A baseline separation of the two enantiomers was obtained at a Cu(II)/L-proline ratio of 1:8 in a borate buffer (15 mmol/L, pH 9.6) containing 10% v/v acetonitrile. Under the optimized conditions, precision linearity range and detection limits of the developed enantioselective CE method were evaluated and compared using two different detection systems: conventional UV detection at 226 nm and iodine (127I)specific detection ("chiral speciation") with ICP-MS. Both methodologies show adequate analytical performance characteristics with detection limits around 0.30 microg/mL for each enantiomer of T4. Finally, a levothroid pharmaceutical formulation sample was successfully analyzed using both developed methods CE-UV and CE-ICP-MS.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrophoresis, Capillary / methods*
  • Hydrogen-Ion Concentration
  • Ligands
  • Mass Spectrometry / methods*
  • Spectrophotometry, Ultraviolet*
  • Stereoisomerism
  • Temperature
  • Thyroxine / chemistry*

Substances

  • Ligands
  • Thyroxine