    Acad Radiol. 2009 Jun;16(6):689-99.

    Reproducibility and validity of lung density measures from cardiac CT Scans--The Multi-Ethnic Study of Atherosclerosis (MESA) Lung Study.

    Source

    Department of Radiology, University of Iowa Carver College of Medicine, Iowa City, USA.

    Abstract

    RATIONALE AND OBJECTIVES:

    Cardiac computed tomographic (CT) scans for the assessment of coronary calcium scores include approximately 70% of the lung volume and may be useful for the quantitative assessment of emphysema. The reproducibility of lung density measures from cardiac computed tomography and their validity compared to lung density measures from full-lung scans is unknown.

    MATERIALS AND METHODS:

    The Multi-Ethnic Study of Atherosclerosis (MESA) performed paired cardiac CT scans for 6814 participants at baseline and at follow-up. The MESA-Lung Study assessed lung density measures in the lung fields of these cardiac scans, counting voxels below -910 HU as moderate-to-severe emphysema-like lung regions. We evaluated: 1) the reproducibility of lung density measures among 120 randomly selected participants; 2) the comparability of measures acquired on electron beam CT (EBCT) and multidetector CT (MDCT) scanners among 10 participants; and 3) the validity of these measures compared to full-lung scans among 42 participants. Limits of agreement were determined using Bland-Altman approaches.

    RESULTS:

    Percent emphysema measures from paired cardiac scans were highly correlated (r = 0.92-0.95) with mean difference of -0.05% (95% limits of agreement: -8.3, 8.4%). Measures from EBCT and MDCT scanners were comparable (mean difference -0.9%; 95% limits of agreement: -5.1, 3.3%). Percent emphysema measures from MDCT cardiac and MDCT full-lung scans were highly correlated (r = 0.93) and demonstrated reasonable agreement (mean difference 2.2%; 95% limits of agreement: -9.2, 13.8%).

    CONCLUSIONS:

    Although full-lung imaging is preferred for the quantification of emphysema, the lung imaging from paired cardiac computed tomography provided a reproducible and valid quantitative assessment of emphysema in a population-based sample.

    PMID:
    19427979
    [PubMed - indexed for MEDLINE]
    PMCID: PMC2943871
    Free PMC Article

    Images from this publication.See all images (5) Free text

    Figure 2
    Figure 1
    Figure 4
    Figure 5
    Figure 3

      Supplemental Content

      Click here to read Click here to read

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk