Format

Send to:

Choose Destination
See comment in PubMed Commons below
J Neural Eng. 2009 Jun;6(3):034001. doi: 10.1088/1741-2560/6/3/034001. Epub 2009 Apr 15.

Feasibility and safety of longitudinal magnetic resonance imaging in a rodent model with intracortical microwire implants.

Author information

  • 1Department of Bioengineering, The Pennsylvania State University, University Park, PA 16802, USA.

Abstract

The purpose of this communication is to investigate (1) the feasibility of carrying out longitudinal magnetic resonance imaging (MRI) studies in animals with implanted microwire electrodes adapted for MRI compatibility, (2) the effect of MRI studies on the quality of neurophysiological recordings, (3) the use of MRI to study the extent and recovery of tissue damage due to electrode insertion and (4) histological tissue damage due to MRI. There was no evidence of chronic neural damage caused by repeated MRI by any of the measures used nor any statistical difference in the quality of the electrophysiological recordings between animals that had undergone MRI scans and those that had not.