Evaluation of a patient-specific Monte Carlo software for CT dosimetry

Radiat Prot Dosimetry. 2009 Feb;133(4):248-55. doi: 10.1093/rpd/ncp051. Epub 2009 Mar 27.

Abstract

The aim was to validate the ImpactMC computed tomography (CT) dosimetry software that allows patient-specific dose determination. Measured values of head- and body-weighted CT dose index (CTDI(w)) were compared with corresponding values derived using ImpactMC software. A physical anthropomorphic phantom simulating the average adult was employed to study the effect of exposure parameters used to produce the input image set on a normalised dose output and the relationship between exposure parameters selected for simulation on the dose output. The difference between CTDI(w) values obtained through measurements and simulations were found to be up to 12.8 and 18.3% for head and body phantoms, respectively. Exposure parameters of the image set used as input were found to have a minor impact on the normalised dose output. Simulations confirmed the expected linear relationship between dose and tube load and the power law relationship between dose and tube potential. Results demonstrate that ImpactMC may be capable of providing reliable CT dose estimates.