    Bioinformatics. 2009 May 15;25(10):1341-3. Epub 2009 Mar 20.

    Automated diagnosis of LC-MS/MS performance.

    Source

    Proteomics and Informatics Services Facility, Research Resources Center, University of Illinois at Chicago, Chicago, IL 60612, USA. huaxu@uic.edu

    Abstract

    SUMMARY: We report a software scheme for automated diagnosis of liquid chromatography tandem mass spectrometry (LC-MS/MS) system performance. The proposed software scheme provides a robust framework for establishing automated diagnosis of LC-MS/MS system performance for a variety of instruments and experiments. This schematic consists of four main software components: (i) data conversion, (ii) peptide identification, (iii) LC retention time analysis and (iv) system performance evaluation. The implementation of a standard approach for assessing LC-MS/MS system performance enables researchers to apply reliable metrics to assess their workflows performance over different batch experiments. Furthermore, the results from system diagnosis can provide feedback to the workflow to stop batch analysis if system performance falls below prescribed thresholds. A basic implementation of the approach based on the MassMatrix database search and LC retention time analysis programs is presented. AVAILABILITY: An open source implementation of the LC-MS/MS system diagnosis software based on the MassMatrix database search program is freely available to non-commercial users and can be downloaded at www.massmatrix.net.

    PMID:
    19304874
    [PubMed - indexed for MEDLINE]
    PMCID: PMC2677746
    Free PMC Article

    Images from this publication.See all images (1) Free text

    Fig. 1.

      Supplemental Content

      Click here to read Click here to read

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk