    Radiology. 2009 Apr;251(1):67-76. Epub 2009 Feb 12.

    Nonalcoholic fatty liver disease: diagnostic and fat-grading accuracy of low-flip-angle multiecho gradient-recalled-echo MR imaging at 1.5 T.

    Source

    Department of Radiology, University of California, San Diego Medical Center, University of California at San Diego, MR3 Laboratory, 408 Dickinson St, San Diego, CA 92103-8226, USA.

    Abstract

    PURPOSE:

    To assess the accuracy of four fat quantification methods at low-flip-angle multiecho gradient-recalled-echo (GRE) magnetic resonance (MR) imaging in nonalcoholic fatty liver disease (NAFLD) by using MR spectroscopy as the reference standard.

    MATERIALS AND METHODS:

    In this institutional review board-approved, HIPAA-compliant prospective study, 110 subjects (29 with biopsy-confirmed NAFLD, 50 overweight and at risk for NAFLD, and 31 healthy volunteers) (mean age, 32.6 years +/- 15.6 [standard deviation]; range, 8-66 years) gave informed consent and underwent MR spectroscopy and GRE MR imaging of the liver. Spectroscopy involved a long repetition time (to suppress T1 effects) and multiple echo times (to estimate T2 effects); the reference fat fraction (FF) was calculated from T2-corrected fat and water spectral peak areas. Imaging involved a low flip angle (to suppress T1 effects) and multiple echo times (to estimate T2* effects); imaging FF was calculated by using four analysis methods of progressive complexity: dual echo, triple echo, multiecho, and multiinterference. All methods except dual echo corrected for T2* effects. The multiinterference method corrected for multiple spectral interference effects of fat. For each method, the accuracy for diagnosis of fatty liver, as defined with a spectroscopic threshold, was assessed by estimating sensitivity and specificity; fat-grading accuracy was assessed by comparing imaging and spectroscopic FF values by using linear regression.

    RESULTS:

    Dual-echo, triple-echo, multiecho, and multiinterference methods had a sensitivity of 0.817, 0.967, 0.950, and 0.983 and a specificity of 1.000, 0.880, 1.000, and 0.880, respectively. On the basis of regression slope and intercept, the multiinterference (slope, 0.98; intercept, 0.91%) method had high fat-grading accuracy without statistically significant error (P > .05). Dual-echo (slope, 0.98; intercept, -2.90%), triple-echo (slope, 0.94; intercept, 1.42%), and multiecho (slope, 0.85; intercept, -0.15%) methods had statistically significant error (P < .05).

    CONCLUSION:

    Relaxation- and interference-corrected fat quantification at low-flip-angle multiecho GRE MR imaging provides high diagnostic and fat-grading accuracy in NAFLD.

    PMID:
    19221054
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC2663579
    Free PMC Article

    Images from this publication.See all images (3) Free text

    Figure 2:
    Figure 1:
    Figure 3:

      Supplemental Content

      Icon for HighWire Press Icon for PubMed Central

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk