Br J Ophthalmol. 2009 May;93(5):634-7. doi: 10.1136/bjo.2008.150276. Epub 2009 Jan 23.

Measurement of total blood flow in the normal human retina using Doppler Fourier-domain optical coherence tomography.

Author information

  • 1Center for Ophthalmic Optics & Lasers, Doheny Eye Institute, Keck School of Medicine, University of Southern California, Los Angeles, CA 90033, USA. yiminwan@usc.edu

Abstract

AIM:

To measure total retinal blood flow in normal human eyes using Doppler Fourier-domain optical coherence tomography (FD-OCT).

METHODS:

10 normal people aged 35 to 69 years were measured for the right eye using Doppler FD-OCT. Double circular scans around the optic nerve heads were used. Four pairs of circular scans that transected all retinal branch vessels were completed in 2 s. Total retinal blood flow was obtained by summing the flows in the branch veins. Measurements from the eight scans were averaged. Veins with diameters >33 microm were taken into account.

RESULTS:

Total retinal blood flow could be measured in eight of 10 subjects: mean (SD) = 45.6 (3.8) microl/min (range 40.8 to 52.9 microl/min). The coefficient of variation for repeated measurements was 10.5%. Measured vein diameters ranged from 33.3 to 155.4 mum. The averaged flow speed was 19.3 (2.9) mm/s, which did not correlate with vessel diameter. There was no significant difference between flows in the superior and inferior retinal hemispheres.

CONCLUSIONS:

Double circular scanning using Doppler FD-OCT is a rapid and reproducible method to measure total retinal blood flow. These flow values are within the range previously established by laser Doppler flowmetry.

PMID:
19168468
[PubMed - indexed for MEDLINE]