Format

Send to

Choose Destination
See comment in PubMed Commons below
Biosens Bioelectron. 1991;6(5):395-400.

A miniature Clark-type oxygen electrode using a polyelectrolyte and its application as a glucose sensor.

Author information

  • 1Fujitsu Laboratories Ltd., Atsugi, Japan.

Abstract

A miniature Clark-type oxygen electrode was fabricated by anisotropically etching silicon. A two-gold-electrode configuration was used and a double-layered gas-permeable membrane was formed directly on the electrolyte, poly(vinyl-4-ethylpyridinium bromide) in the sensitive area. These materials improved the electrode's stability in long-term storage and sterilization tolerance to a practical level. The 90% response time averaged 80 s and residual current 10%, with a good linear calibration curve. The oxygen electrode was also used to make an integrated sensor for the simultaneous determination of glucose and oxygen. The glucose sensor's response time was 50-110 s, with good linearity in glucose concentrations between 56 microM and 1.1 mM at 37 degrees C, pH 7.0.

PMID:
1910664
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science
    Loading ...
    Write to the Help Desk