    J Biomed Opt. 2008 Sep-Oct;13(5):054019.

    Double-layer estimation of intra- and extracerebral hemoglobin concentration with a time-resolved system.

    Source

    Institut de Génie Biomédical, Ecole Polytechnique de Montréal, C.P. 6079, succ. Centre-Ville, Montréal, Québec H3C 3A7, Canada. louis-2.gagnon@polymtl.ca

    Abstract

    We present in vivo measurements of baseline physiology from five subjects with a four-wavelength (690, 750, 800, and 850 nm) time-resolved optical system. The measurements were taken at four distances: 10, 15, 25, and 30 mm. All distances were fit simultaneously with a two-layered analytical model for the absorption and reduced scattering coefficient of both layers. The thickness of the first layer, comprising the skin, scalp, and cerebrospinal fluid, was obtained from anatomical magnetic resonance images. The fitting procedure was first tested with simulations before being applied to in vivo measurements and verified that this procedure permits accurate characterization of the hemoglobin concentrations in the extra- and intracerebral tissues. Baseline oxyhemoglobin, deoxyhemoglobin, and total hemoglobin concentrations and oxygen saturation were recovered from in vivo measurements and compared to the literature. We observed a noticeable intersubject variability of the hemoglobin concentrations, but constant values for the cerebral hemoglobin oxygen saturation.

    PMID:
    19021399
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC2718835
    Free PMC Article

    Images from this publication.See all images (3) Free text

    Fig. 1
    Fig. 3
    Fig. 2

      Supplemental Content

      Icon for American Institute of Physics Icon for PubMed Central

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk