Display Settings:

Format

Send to:

Choose Destination
    Am J Phys Med Rehabil. 2009 Apr;88(4):264-6.

    Improper use of a transtibial prosthesis silicone liner causing pressure ulceration.

    Source

    Division of Physical Medicine and Rehabilitation, Department of Medicine, Dalhousie University, QEII, Nova Scotia Rehabilitation Centre, Halifax, Nova Scotia, Canada.

    Abstract

    We present the previously unreported hazard of creating pressure ulceration in a susceptible host by the improper use of a silicone prosthetic liner. An 80-yr-old man sustained a recent transtibial amputation for peripheral vascular disease. His comorbidities included vascular cognitive impairment, type 2 diabetes mellitus, coronary artery disease, anemia of chronic disease, postherpetic neuralgia, and pruritus of uncertain origin. When not using his transtibial prosthesis, he found his 1.5-mm thick silicone liner (ICEROSS) more comfortable to wear than his stump shrinker and thermoplastic protector. Moreover, he repeatedly wore his liner rolled partway down his stump to allow him to scratch pruritic skin. A horizontal, linear, stage 2 ulcer developed on the residual limb under the upper edge of the rolled liner. The ulcer required >3 mos to heal. This case illustrates the importance of candidate selection for roll-on liners, proper patient and professional training, and optimal management of patient comorbidities.

    PMID:
    18971772
    [PubMed - indexed for MEDLINE]

      Supplemental Content

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk