Abstract

A new, simple, clean and selective flow injection strategy based on the tandem photochemical reaction-chemiluminescence detection was applied to the determination of chloramphenicol. The determination is based on the on-line photodegradation of the drug in a glycine buffer at pH 8.8 by using a photoreactor consisting of 697 cmx0.5 mm PTFE tubing helically coiled around an 8 W low-pressure mercury lamp. Photodegradated chloramphenicol is detected by direct chemiluminescence of resulting photo-fragments and their subsequent reaction with potassium permanganate in sulfuric acid medium as oxidant. The method allows the chemiluminescence determination of compounds which do not exhibit native chemiluminescence. The calibration graph was linear up to 14 mug ml(-1) chloramphenicol, the limit of detection was 30 ng ml(-1), the relative standard deviation was 2.4% for 7 mug ml(-1) of the drug and the sample throughput was 60 h(-1). Taking into account the importance of the medium of photodegradation on the mechanism of photodegradation a comparative study in terms of selective was performed for different chemical media employed in the procedure of photodegradation. The proposed method was applied to the determination of chloramphenicol in commercially available pharmaceutical formulations.