There are currently no ideal quantitative measures assessing osteolysis after total joint arthroplasty. This study reports a new approach for quantification of osteolytic lesions on radiographs of total hip arthroplasty with the hypothesis that computer-assisted lesion quantification improves analysis precision over currently used qualitative visual assessment and thus improves progression monitoring. Duplicate exposure sets of anterior-posterior and frog-leg view radiographs of 15 volunteer total hip arthroplasty patients with radiographically evident periprosthetic lucencies were used. Two independent readers delineated the lesions using on-screen digital marker tools. Based on duplicate readings of the first exposure set only, intraoperator precision had a coefficient of variation (CV) from 1.5% to 3.4%, whereas interoperator precision CV ranged from 3.2% to 4.6%. The reproducibility of the x-ray technique as assessed by single readings of both sets of x-rays was CV 3.2% to 4.8%. The new technique compares favorably to precision of qualitative visual assessment and permits more accurate detection and quantitation of osteolytic lesions.

PMID:
18534519
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science
    Loading ...
    Write to the Help Desk