    Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Jul;106(1):106-14. Epub 2008 May 27.

    Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology.

    Source

    Department of Diagnostic Sciences and General Dentistry, University of North Carolina School of Dentistry, Chapel Hill, North Carolina 27599-7450, USA. jbl@email.unc.edu

    Abstract

    OBJECTIVES:

    This study compares 2 measures of effective dose, E(1990) and E(2007), for 8 dentoalveolar and maxillofacial cone-beam computerized tomography (CBCT) units and a 64-slice multidetector CT (MDCT) unit.

    STUDY DESIGN:

    Average tissue-absorbed dose, equivalent dose, and effective dose were calculated using thermoluminescent dosimeter chips in a radiation analog dosimetry phantom. Effective doses were derived using 1990 and the superseding 2007 International Commission on Radiological Protection (ICRP) recommendations.

    RESULTS:

    Large-field of view (FOV) CBCT E(2007) ranged from 68 to 1,073 microSv. Medium-FOV CBCT E(2007) ranged from 69 to 560 microSv, whereas a similar-FOV MDCT produced 860 microSv. The E(2007) calculations were 23% to 224% greater than E(1990).

    CONCLUSIONS:

    The 2007 recommendations of the ICRP, which include salivary glands, extrathoracic region, and oral mucosa in the calculation of effective dose, result in an upward reassessment of fatal cancer risk from oral and maxillofacial radiographic examinations. Dental CBCT can be recommended as a dose-sparing technique in comparison with alternative medical CT scans for common oral and maxillofacial radiographic imaging tasks.