Display Settings:

Format

Send to:

Choose Destination
See comment in PubMed Commons below
J Thromb Haemost. 2008 Aug;6(8):1304-12. doi: 10.1111/j.1538-7836.2008.03025.x. Epub 2008 May 17.

Quantitative interpretation of optical density measurements using PF4-dependent enzyme-immunoassays.

Author information

  • 1Department of Pathology and Molecular Medicine, Michael G DeGroote School of Medicine, McMaster University, Hamilton, ON, Canada. twarken@mcmaster.ca

Abstract

BACKGROUND:

Many laboratories test for heparin-induced thrombocytopenia (HIT) using a PF4-dependent enzyme-immunoassay (EIA). An advantage of the EIA is its simplicity; a disadvantage is that it only indirectly detects heparin-dependent, platelet-activating antibodies ('HIT antibodies').

OBJECTIVES:

To determine whether the magnitude of a positive EIA result, expressed in optical density (OD) units, predicts risk of HIT antibodies, defined as a strong-positive platelet serotonin-release assay (SRA) result (>or=50% serotonin release).

PATIENTS/METHODS:

We determined the risk of a strong-positive SRA result for five categories of OD reactivity (<0.40, 0.40-<1.00, 1.00-<1.40, 1.40-<2.00, and >or=2.00 OD units) using two EIAs (commercial anti-PF4/polyanion IgG/A/M and in-house anti-PF4/heparin-IgG).

RESULTS:

For patient sera investigated for HIT antibodies, a weak-positive result (0.40-<1.00 OD units) in either EIA indicated a low probability (<or=5%) of a strong-positive SRA; the risk increased to approximately 90% with an OD >or= 2.00 units. Quantifying the EIA-SRA relationship for 1553 referred patient sera, we found that for every increase of 0.50 OD units in the EIA-IgG, the risk of a strong-positive SRA result increased by OR = 6.39 [95% confidence interval (CI), 5.13, 7.95; P < 0.0001]. For every increase of 1.00 OD units in the EIA-IgG, the risk increased by OR = 40.81 (95% CI, 26.35, 63.20; P < 0.0001).

CONCLUSIONS:

The probability of HIT antibodies (strong-positive SRA result) inferred by a positive PF4-dependent EIA varies considerably in relation to the magnitude of the EIA result, expressed as OD values. In our laboratory, the probability of HIT antibodies being present reached >or=50% only when the OD level was >or=1.40 units.

Comment in

PMID:
18489711
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Blackwell Publishing
    Loading ...
    Write to the Help Desk