Isoform separation of a multi-acetylated protein using capillary polystyrene-divinylbenzene monolithic columns.

Dionex Corporation, Amsterdam, The Netherlands. Evert-Jan.Sneekes@dionex.com

Capillary polystyrene-divinylbenzene (PS-DVB) monolithic columns were used to separate differentially acetylated intact IM9 protein isoforms. Compared to the unmodified form, the hydrophobic shift for intact acetylated isoforms was significant under standard reversed-phase conditions (32.5-45% acetonitrile in 10 min). The high chromatographic resolution of the PS-DVB monolithic columns resulted in peak widths at half height of 4-5s. This allowed us to nearly completely resolve a number of peaks greater than the number of possible acetylation sites. This observation suggested that not only the number, but also the location of the acetylations on the protein had a significant effect on the retention. Matrix-assisted laser desorption ionization time-of-flight MS and MS/MS were used to confirm the chromatographic separation of isoforms. It was found that the acetylations site, especially on the N-terminus, has an effect on the retention on the PS-DVB column.