Local re-inversion coronary MR angiography: arterial spin-labeling without the need for subtraction

J Magn Reson Imaging. 2008 Apr;27(4):913-7. doi: 10.1002/jmri.21319.

Abstract

Purpose: To implement a double-inversion bright-blood coronary MR angiography sequence using a cylindrical re-inversion prepulse for selective visualization of the coronary arteries.

Materials and methods: Local re-inversion bright-blood magnetization preparation was implemented using a nonselective inversion followed by a cylindrical aortic re-inversion prepulse. After an inversion delay that allows for in-flow of the labeled blood-pool into the coronary arteries, three-dimensional radial steady-state free-precession (SSFP) imaging (repetition/echo time, 7.2/3.6 ms; flip angle, 120 degrees, 16 profiles per RR interval; field of view, 360 mm; matrix, 512, twelve 3-mm slices) is performed. Coronary MR angiography was performed in three healthy volunteers and in one patient on a commercial 1.5 Tesla whole-body MR System.

Results: In all subjects, coronary arteries were selectively visualized with positive contrast. In addition, a middle-grade stenosis of the proximal right coronary artery was seen in one patient.

Conclusion: A novel T1 contrast-enhancement strategy is presented for selective visualization of the coronary arteries without extrinsic contrast medium application. In comparison to former arterial spin-labeling schemes, the proposed magnetization preparation obviates the need for a second data set and subtraction.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Coronary Circulation
  • Coronary Stenosis / diagnosis
  • Coronary Vessels / anatomy & histology*
  • Humans
  • Magnetic Resonance Angiography*
  • Middle Aged
  • Spin Labels*
  • Subtraction Technique

Substances

  • Spin Labels