J Thorac Cardiovasc Surg. 2008 Jan;135(1):196-202. doi: 10.1016/j.jtcvs.2007.08.043.

Effects of visual force feedback on robot-assisted surgical task performance.

Author information

  • 1Engineering Research Center and Computer Integrated Surgery Systems and Technology, Johns Hopkins University, Baltimore, MD, USA.

Abstract

OBJECTIVE:

Direct haptic (force or tactile) feedback is negligible in current surgical robotic systems. The relevance of haptic feedback in robot-assisted performances of surgical tasks is controversial. We studied the effects of visual force feedback, a haptic feedback surrogate, on tying surgical knots with fine sutures similar to those used in cardiovascular surgery.

METHODS:

By using a modified da Vinci robotic system (Intuitive Surgical, Inc, Sunnyvale, Calif) equipped with force-sensing instrument tips and real-time visual force feedback overlays in the console image, 10 surgeons each tied 10 knots with and 10 knots without visual force feedback. Four surgeons had significant prior da Vinci experience, and the remaining 6 surgeons did not. Performance parameters, including suture breakage and secure knots, peak and standard deviation of applied forces, and completion times using 5-0 silk sutures, were recorded. Chi-square and Student t test analyses determined the differences between groups.

RESULTS:

Among surgeon subjects with robotic experience, no differences in measured performance parameters were found between robot-assisted knot ties executed with and without visual force feedback. Among surgeons without robotic experience, however, visual force feedback was associated with lower suture breakage rates, peak applied forces, and standard deviations of applied forces. Visual force feedback did not impart differences in knot completion times or loose knots for either surgeon group.

CONCLUSIONS:

Visual force feedback resulted in reduced suture breakage, lower forces, and decreased force inconsistencies among novice robotic surgeons, although elapsed time and knot quality were unaffected. In contrast, visual force feedback did not affect these metrics among surgeons experienced with the da Vinci system. These results suggest that visual force feedback primarily benefits novice robot-assisted surgeons, with diminishing benefits among experienced surgeons.

PMID:
18179942
[PubMed - indexed for MEDLINE]
PMCID:
PMC2674617
Free PMC Article

Images from this publication.See all images (4)Free text

Fig 1
Fig 2
Fig 3
Fig 4
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science Icon for PubMed Central
    Loading ...
    Write to the Help Desk