1: J Am Chem Soc. 2008 Jan 23;130(3):794-5. Epub 2007 Dec 23.Click here to read Links

Paramagnetic relaxation-based 19f MRI probe to detect protease activity.

Division of Advanced Science and Biotechnology, Graduate School of Engineering, Osaka University, Osaka 565-0871, Japan.

A novel design principle for 19F MRI probes detecting protease activity was developed. This principle is based on 19F MRI signal quenching by the intramolecular paramagnetic effect from Gd3+. The intramolecular Gd3+ dramatically attenuated the 19F probe signal, and the paramagnetic effect was cancelled by the probe hydrolyzation by caspase-3. Using this probe, it was shown that the probe could detect caspase-3 activity spatially from a phantom image using 19F MRI.

PMID: 18154336 [PubMed - indexed for MEDLINE]