Display Settings:

Format

Send to:

Choose Destination
Radiology. 2008 Feb;246(2):581-8. Epub 2007 Nov 30.

Automated quantification of myocardial infarction from MR images by accounting for partial volume effects: animal, phantom, and human study.

Author information

  • 1Department of Radiology, Lund University Hospital, SE-221 85 Lund, Sweden.

Abstract

Ethics committees approved human and animal study components; informed written consent was provided (prospective human study [20 men; mean age, 62 years]) or waived (retrospective human study [16 men, four women; mean age, 59 years]). The purpose of this study was to prospectively evaluate a clinically applicable method, accounting for the partial volume effect, to automatically quantify myocardial infarction from delayed contrast material-enhanced magnetic resonance images. Pixels were weighted according to signal intensity to calculate infarct fraction for each pixel. Mean bias +/- variability (or standard deviation), expressed as percentage left ventricular myocardium (%LVM), were -0.3 +/- 1.3 (animals), -1.2 +/- 1.7 (phantoms), and 0.3 +/- 2.7 (patients), respectively. Algorithm had lower variability than dichotomous approach (2.7 vs 7.7 %LVM, P < .01) and did not differ from interobserver variability for bias (P = .31) or variability (P = .38). The weighted approach provides automatic quantification of myocardial infarction with higher accuracy and lower variability than a dichotomous algorithm.

(c) RSNA, 2007.

PMID:
18055873
[PubMed - indexed for MEDLINE]