My NCBISign In

Display Settings:

Format

Send to:

Choose Destination

    Theor Biol Med Model. 2007 Aug 23;4:31.

    Variance in multiplex suspension array assays: microsphere size variation impact.

    Hanley BP, Xing L, Cheng RH.

    Microbiology Graduate Group, University of California, Davis, CA 95616, USA. bphanley@ucdavis.edu

    BACKGROUND: Luminex suspension microarray assays are in widespread use. There are issues of variability of assay readings using this technology. METHODS AND RESULTS: Size variation is demonstrated by transmission electron microscopy. Size variations of microspheres are shown to occur in stepwise increments. A strong correspondence between microsphere size distribution and distribution of fluorescent events from assays is shown. An estimate is made of contribution of microsphere size variation to assay variance. CONCLUSION: A probable significant cause of variance in suspended microsphere assay results is variation in microsphere diameter. This can potentially be addressed by changes in the manufacturing process. Provision to users of mean size, median size, skew, the number of standard deviations that half the size range represents (sigma multiple), and standard deviation is recommended. Establishing a higher sigma multiple for microsphere production is likely to deliver a significant improvement in precision of raw instrument readings. Further research is recommended on the molecular architecture of microsphere coatings.

    PMID: 17716372 [PubMed - indexed for MEDLINE]