Reactive desorption electrospray ionization for selective detection of the hydrolysis products of phosphonate esters

J Mass Spectrom. 2007 Aug;42(8):1086-92. doi: 10.1002/jms.1244.

Abstract

Reactive desorption electrospray ionization (reactive DESI) is demonstrated to be a rapid and sensitive method for the direct detection of alkyl methylphosphonic acids, the hydrolysis products and metabolites of the chemical warfare (CW) agents VX (S-2-diisopropylaminoethyl-O-ethyl methylphosphonothiolate) and GB (sarin, isopropylmethyl phosphonofluoridate). Rapid and sensitive detection of these compounds is readily achieved by performing DESI from a solid surface; detection specificity is enhanced by implementation of a heterogeneous ion/molecule reaction using boric acid in the spray solvent. The reagent ion H(2)BO(3) (-) generated in the spray readily reacts with condensed-phase alkyl MPA to form anionic adducts. The specificity of this chemical reaction, together with the characteristic fragmentation patterns of the reaction products, supplies a highly discriminatory detection method for methylphosphonic acid (MPA), ethylphosphonic acid (EMPA) and isopropyl methylphosphonic acid (IMPA) in complex matrices.