    J Clin Microbiol. 2007 Aug;45(8):2529-36. Epub 2007 Jun 13.

    Comparison of performance characteristics of three real-time reverse transcription-PCR test systems for detection and quantification of hepatitis C virus.

    Source

    Molecular Diagnostics Laboratory, Department of Pathology, Virginia Commonwealth University, Richmond, Virginia 23298-0248, USA.

    Abstract

    We evaluated the performance characteristics of three real-time reverse transcription-PCR test systems for detection and quantification of hepatitis C virus (HCV) and performed a direct comparison of the systems on the same clinical specimens. Commercial HCV panels (genotype 1b) were used to evaluate linear range, sensitivity, and precision. The Roche COBAS TaqMan HCV test for research use only (RUO) with samples processed on the MagNA Pure LC instrument (Roche RUO-MPLC) and Abbott analyte-specific reagents (ASR) with QIAGEN sample processing (Abbott ASR-Q) showed a sensitivity of 1.0 log(10) IU/ml with a linear dynamic range of 1.0 to 7.0 log(10) IU/ml. The Roche ASR in combination with the High Pure system (Roche ASR-HP) showed a sensitivity of 1.4 log(10) IU/ml with a linear dynamic range of 2.0 to 7.0 log(10) IU/ml. All of the systems showed acceptable reproducibility, the Abbott ASR-Q being the most reproducible of the three systems. Seventy-six clinical specimens (50 with detectable levels of HCV RNA and various titers and genotypes) were tested, and results were compared to those of the COBAS Amplicor HCV Monitor v2.0. Good correlation was obtained for the Roche RUO-MPLC and Abbott ASR-Q (R(2) = 0.84 and R(2) = 0.93, respectively), with better agreement for the Abbott ASR-Q. However, correlation (R(2) = 0.79) and agreement were poor for Roche ASR-HP, with bias relative to concentration and genotype. Roche ASR-HP underestimated HCV RNA for genotypes 3 and 4 as much as 2.19 log(10) IU/ml. Our study demonstrates that Roche RUO-MPLC and Abbott ASR-Q provided acceptable results and agreed sufficiently with the COBAS Amplicor HCV Monitor v2.0.

    PMID:
    17567786
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC1951251
    Free PMC Article

    Images from this publication.See all images (4)Free text

    FIG. 1.
    FIG. 2.
    FIG. 3.
    FIG. 4.

      Supplemental Content

      Icon for HighWire Icon for PubMed Central

      Save items

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk