Display Settings:

Format

Send to:

Choose Destination
See comment in PubMed Commons below
Arch Dis Child Fetal Neonatal Ed. 2007 May;92(3):F219-24.

Should pulse oximetry be used to screen for congenital heart disease?

Author information

  • 1Department of Paediatrics, PO Box 8041, 96101 Rovaniemi, Finland. pekka.valmari@lshp.fi

Abstract

Ten studies (44 969 newborns, 71 severe defects) evaluating the usefulness of neonatal pulse oximetry (PO) screening in timely detection of congenital heart disease (CHD) were reviewed. PO showed a high specificity (99.9-99.99%), and the overall rate of detection of 15 individual defects with PO was 72% (range 46-100%), exceeding that of the clinical examination 58% (9-86%). Similar results were obtained for cyanotic CHD (89% v 69%, respectively). Without PO, discharge of apparently healthy infants with unknown CHD was 5.5 times and 4.1 times more likely in cyanotic CHD and all serious CHD, respectively. The paper describes the technical and practical details of first day and later screening. Diagnosis is reached earliest with first day screening, but it requires more resources. PO screening is not sensitive enough to serve as an independent screen, but along with the clinical examination it helps minimise the morbidity and mortality associated with discharge without diagnosis. Further research is needed for precise delineation of populations that would benefit from PO screening.

PMID:
17449857
[PubMed - indexed for MEDLINE]
PMCID:
PMC2675337
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Icon for HighWire Icon for PubMed Central
    Loading ...
    Write to the Help Desk