J Opt Soc Am A Opt Image Sci Vis. 2007 May;24(5):1327-36.

Compact multimodal adaptive-optics spectral-domain optical coherence tomography instrument for retinal imaging.

Author information

  • 1Physical Sciences Inc., Andover, MA 01810, USA.

Abstract

We have developed a compact, multimodal instrument for simultaneous acquisition of en face quasi-confocal fundus images and adaptive-optics (AO) spectral-domain optical coherence tomography (SDOCT) cross-sectional images. The optical system including all AO and SDOCT components occupies a 60x60 cm breadboard that can be readily transported for clinical applications. The AO component combines a Hartmann-Shack wavefront sensor and a microelectromechanical systems-based deformable mirror to sense and correct ocular aberrations at 15 Hz with a maximum stroke of 4 microm. A broadband superluminescent diode source provides 4 mum depth resolution for SDOCT imaging. In human volunteer testing, we observed up to an 8 dB increase in OCT signal and a corresponding lateral resolution of <10 microm as a result of AO correction.