Coupling chemiluminescence with capillary electrophoresis to analyze single human red blood cells

Anal Chim Acta. 2007 Feb 5;583(2):217-22. doi: 10.1016/j.aca.2006.09.068. Epub 2006 Oct 21.

Abstract

In this paper, we describe a new method for determination of hemoglobin of single red blood cells by coupling chemiluminescence with capillary electrophoresis (CL-CE). The chemiluminescent detection is based on the catalytic effects of hemoglobin on the luminol-hydrogen peroxide reaction. The conditions of chemiluminescent reaction and capillary electrophoresis were investigated. Hemoglobin in human blood samples was detected with the present method, the linear range from 1.7 microg mL(-1) to 6.8 microg mL(-1) was tested, and the correlation coefficient of 0.997 and low detection limit of 0.17 microg mL(-1) (approximately 2.2 pg, S/N=3) were obtained. Cell injection procedure was improved, and the method was successfully used to determine hemoglobin of single red blood cells and the statistical result of the average content of hemoglobin in 26 human red blood cells was 23.6 pg. Compared to other current methods, CE with CL system is simple, sensitive and will become an attractive alternative method for single cell analysis.