Display Settings:

Format

Send to:

Choose Destination
    Annu Rev Biomed Eng. 2007;9:389-413.

    SQUID-detected magnetic resonance imaging in microtesla fields.

    Source

    Department of Physics, University of California, Berkeley, California 94720-7300, USA. jclarke@berkeley.edu

    Abstract

    The use of very low noise magnetometers based on Superconducting QUantum Interference Devices (SQUIDs) enables nuclear magnetic resonance (NMR) and magnetic resonance imaging (MRI) in microtesla magnetic fields. An untuned superconducting flux transformer coupled to a SQUID achieves a magnetic field noise of 10(-15) T Hz(-1/2). The frequency-independent response of this magnetometer combined with prepolarization of the nuclear spins yields an NMR signal that is independent of the Larmor frequency omega0. An MRI system operating in a field of 132 microT, corresponding to a proton frequency of 5.6 kHz, achieves an in-plane resolution of 0.7 x 0.7 mm2 in phantoms. Measurements of the longitudinal relaxation time T1 in different concentrations of agarose gel over five decades of frequency reveal much greater T1-differentiation at fields below a few millitesla. Microtesla MRI has the potential to image tumors with substantially greater T1-weighted contrast than is achievable in high fields in the absence of a contrast agent.