Display Settings:

Format

Send to:

Choose Destination
See comment in PubMed Commons below
Phys Rev Lett. 2006 Sep 29;97(13):130801. Epub 2006 Sep 26.

Accurate optical lattice clock with 87Sr atoms.

Author information

  • 1LNE-SYRTE, Observatoire de Paris 61, Avenue de l'Observatoire, 75014 Paris, France.

Abstract

We report a frequency measurement of the 1S0-3P0 transition of 87Sr atoms in an optical lattice clock. The frequency is determined to be 429 228 004 229 879(5) Hz with a fractional uncertainty that is comparable to state-of-the-art optical clocks with neutral atoms in free fall. The two previous measurements of this transition were found to disagree by about 2 x 10(-13), i.e., almost 4 times the combined error bar and 4 to 5 orders of magnitude larger than the claimed ultimate accuracy of this new type of clocks. Our measurement is in agreement with one of these two values and essentially resolves this discrepancy.