    IEEE Trans Med Imaging. 2006 Oct;25(10):1329-40.

    Prostate mechanical imaging: 3-D image composition and feature calculations.

    Source

    Artann Laboratories, Inc., Trenton, NJ 08618, USA.

    Abstract

    We have developed a method and a device entitled prostate mechanical imager (PMI) for the real-time imaging of prostate using a transrectal probe equipped with a pressure sensor array and position tracking sensor. PMI operation is based on measurement of the stress pattern on the rectal wall when the probe is pressed against the prostate. Temporal and spatial changes in the stress pattern provide information on the elastic structure of the gland and allow two-dimensional (2-D) and three-dimensional (3-D) reconstruction of prostate anatomy and assessment of prostate mechanical properties. The data acquired allow the calculation of prostate features such as size, shape, nodularity, consistency/hardness, and mobility. The PMI prototype has been validated in laboratory experiments on prostate phantoms and in a clinical study. The results obtained on model systems and in vivo images from patients prove that PMI has potential to become a diagnostic tool that could largely supplant DRE through its higher sensitivity, quantitative record storage, ease-of-use and inherent low cost.

    PMID:
    17024836
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC2572682
    Free PMC Article

    Images from this publication.See all images (8) Free text

    Fig. 1
    Fig. 3
    Fig. 5
    Fig. 7
    Fig. 2
    Fig. 4
    Fig. 6
    Fig. 8

      Supplemental Content

      Icon for PubMed Central

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk