    JSLS. 2006 Apr-Jun;10(2):226-30.

    Comparison of cervical detachment using monopolar lap loop ligature and conventional methods in laparoscopic supracervical hysterectomy.

    Source

    Department of Obstetrics, Gynecology & Women's Health, University of Louisville School of Medicine, Louisville, Kentucky 40292, USA. paya@louisville.edu

    Abstract

    OBJECTIVES:

    Our aim was to evaluate a new electrosurgical instrument (Lap Loop device) that amputates the uterine corpus from the cervix during a laparoscopic supracervical hysterectomy (LSH) and to compare the time required for cervical amputation with traditional methods.

    METHODS:

    This comparative trial was conducted at the University of Louisville and Norton Healthcare Hospitals, Louisville, KY. The patients comprised 29 women scheduled for hysterectomy for benign conditions. All patients underwent laparoscopic supracervical hysterectomy. The Lap Loop device was used in 17 patients to section the cervix. Conventional methods with either laparoscopic monopolar scissors or Harmonic scalpel were used in 12 patients.

    RESULTS:

    In the control group, the mean cervical cutting time with laparoscopic scissors or Harmonic scalpel was 14.4 minutes. The mean time for the application of the loop electrode and cutting time was 6.6 minutes and was significantly shorter than the cutting time of conventional methods. Two minor complications (7.4% of cases) and one relatively major complication (3.7%), an incisional hernia, occurred in the study patients. None of the complications were related to the new device.

    CONCLUSION:

    An electrosurgical loop decreased the time required for resection of the uterine cervix during LSH for benign uterine conditions. This device facilitates and increases the safety of this procedure.

    PMID:
    16882425
    [PubMed - indexed for MEDLINE]
    PMCID: PMC3016112
    Free PMC Article

    Images from this publication.See all images (2) Free text

    Figure 1.
    Figure 2.

      Supplemental Content

      Click here to read Click here to read

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk