Display Settings:

Format

Send to:

Choose Destination
    Breast J. 2006 Jul-Aug;12(4):309-23.

    High-resolution fluorodeoxyglucose positron emission tomography with compression ("positron emission mammography") is highly accurate in depicting primary breast cancer.

    Source

    American Radiology Services, Johns Hopkins Green Spring, Lutherville, Maryland 21093, USA. wendieberg@hotmail.com

    Abstract

    We sought to prospectively assess the diagnostic performance of a high-resolution positron emission tomography (PET) scanner using mild breast compression (positron emission mammography [PEM]). Data were collected on concomitant medical conditions to assess potential confounding factors. At four centers, 94 consecutive women with known breast cancer or suspicious breast lesions received 18F-fluorodeoxyglucose (FDG) intravenously, followed by PEM scans. Readers were provided clinical histories and x-ray mammograms (when available). After excluding inevaluable cases and two cases of lymphoma, PEM readings were correlated with histopathology for 92 lesions in 77 women: 77 index lesions (42 malignant), 3 ipsilateral lesions (3 malignant), and 12 contralateral lesions (3 malignant). Of 48 cancers, 16 (33%) were clinically evident; 11 (23%) were ductal carcinoma in situ (DCIS), and 37 (77%) were invasive (30 ductal, 4 lobular, and 3 mixed; median size 21 mm). PEM depicted 10 of 11 (91%) DCIS and 33 of 37 (89%) invasive cancers. PEM was positive in 1 of 2 T1a tumors, 4 of 6 T1b tumors, 7 of 7 T1c tumors, and 4 of 4 cases where tumor size was not available (e.g., no surgical follow-up). PEM sensitivity for detecting cancer was 90%, specificity 86%, positive predictive value (PPV) 88%, negative predictive value (NPV) 88%, accuracy 88%, and area under the receiver-operating characteristic curve (Az) 0.918. In three patients, cancer foci were identified only on PEM, significantly changing patient management. Excluding eight diabetic subjects and eight subjects whose lesions were characterized as clearly benign with conventional imaging, PEM sensitivity was 91%, specificity 93%, PPV 95%, NPV 88%, accuracy 92%, and Az 0.949 when interpreted with mammographic and clinical findings. FDG PEM has high diagnostic accuracy for breast lesions, including DCIS.

    PMID:
    16848840
    [PubMed - indexed for MEDLINE]

      Supplemental Content

      Click here to read

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk