Display Settings:

Format

Send to:

Choose Destination
We are sorry, but NCBI web applications do not support your browser and may not function properly. More information
J Pharm Biomed Anal. 2006 Oct 11;42(4):494-9. Epub 2006 Jun 9.

Enhancement of an analytical method for the determination of squalene in anthrax vaccine adsorbed formulations.

Author information

  • 1SRI International, Menlo Park, CA 94025, USA. ronald.spanggord@sri.com

Abstract

Specific lots of anthrax vaccine adsorbed administered to members of the U.S. Armed Forces have been alleged to contain squalene, a chemical purported to be associated with illnesses of Gulf War veterans. A method of enhanced sensitivity for determining squalene in anthrax vaccine adsorbed using high-performance liquid chromatography with photodiode array detection has been developed, validated, and applied to 44 bottles of 38 lots of anthrax vaccine. In 43 bottles of 37 lots, no squalene was detected within a detection limit of 1ng/0.5ml dose (2 parts-per-billion). One lot, FAV008, was found to contain trace amounts of squalene at 7, 9, and 1microgl(-1), levels considerably below normal human plasma levels (290microgl(-1)). The overall results of this investigation provide direct evidence for the absence of squalene in nearly all of anthrax vaccine preparations tested.

PMID:
16762524
[PubMed - indexed for MEDLINE]