Opt Lett. 2006 May 15;31(10):1477-9.

Resolution enhancement in a light-sheet-based microscope (SPIM).

Author information

  • 1Light Micrscopy Group, EMBL Heidelberg, Germany.

Abstract

Light-sheet-based microscopy [single-plane illumination microscope (SPIM)] performs very well at low numerical apertures. It complements conventional (FM), confocal (CFM), and two-photon fluorescence microscopy (2hnu-FM) currently used in modern life sciences. Lateral and axial SPIM point spread function (PSF) extents are measured by using fluorescent beads to determine the 3D resolution. The results are compared with values derived from an analytical theory and numerical simulations. The discrepancies are found to be less than 5%. The axial extent of a SPIM-PSF (10x/0.3 W) is approximately 5.7 microm. This value is almost a factor of 2 smaller than in CFM, more than 2.5 times smaller than in FM, and more than three times smaller than in 2hnu-FM. SPIM outperforms 2hnu-FM and FM, while CFM has a better axial resolution at NAs above 0.8.

PMID:
16642144
[PubMed - indexed for MEDLINE]