Abstract
STUDY DESIGN:
Prospective cohort study of 52 patients who had undergone artificial lumbar disc replacement.
OBJECTIVES:
To evaluate the implantation accuracy of prosthesis positioning, subsequent facet joint changes and prosthesis migration, and the clinical consequences of implant position.
SUMMARY OF BACKGROUND DATA:
Accuracy of spinal prosthesis implantation has not been evaluated rigorously, especially with a mini-incision approach. It is unknown if the inexact placement of a mobile device in the spine has any biomechanical, radiographic, or clinical repercussions.
METHODS:
A total of 52 consecutive patients were treated using standard methods of disc implantation with an intervertebral prosthesis. Computed tomography scans were performed within 3 days and again at 6 to 24 months. An independent radiologist analyzed the images for prosthesis position, rotation, migration, and facet changes. Results were compared with clinical outcome, measured by the Visual Analog Scale and Oswestry Disability Index.
RESULTS:
Deviation of the prosthesis from the center position was under 1.2 mm, and rotation off of midline was under 12 degrees. Follow-up CT scans showed no migration or facet changes. Regression analysis showed no correlation of prosthesis position with clinical outcome.
CONCLUSIONS:
Current prosthetic disc implantation methods, with minimally invasive access techniques, are relatively accurate. Although there can be deviation of the prosthesis from ideal placement, no repercussions were attributable.