Display Settings:

Format

Send to:

Choose Destination
    Clin Orthop Relat Res. 1991 Nov;(272):255-8.

    Histologic analysis of a retrieved hydroxyapatite-coated femoral prosthesis.

    Source

    Department of Orthopaedic Surgery, Hilleroed Hospital, Denmark.

    Abstract

    A hydroxyapatite-coated hip hemi-prosthesis was retrieved from a 98-year-old osteoporotic woman 12 weeks after implantation. Histologic analysis revealed bone and fibrous tissue almost evenly distributed around the surface of the implant circumference. Quantitative histologic analysis showed that 48% of the hydroxyapatite surface was covered by bone. Fibrous tissue covered 30% of the prosthetic surface, and 20% of the surface had no tissue coverage. Scanning electron microscopy showed direct contact without any clear boundary between the newly formed bone and the hydroxyapatite ceramic.

    PMID:
    1657474
    [PubMed - indexed for MEDLINE]