Display Settings:

Format

Send to:

Choose Destination
    Biomed Eng Online. 2006 Mar 24;5:22.

    Antimicrobial titanium/silver PVD coatings on titanium.

    Source

    Department for Functional Materials in Medicine and Dentistry, University of Würzburg, Pleicherwall 2, D-97070 Würzburg, Germany. andrea.ewald@fmz.uni-wuerzburg.de

    Abstract

    BACKGROUND:

    Biofilm formation and deep infection of endoprostheses is a recurrent complication in implant surgery. Post-operative infections may be overcome by adjusting antimicrobial properties of the implant surface prior to implantation. In this work we described the development of an antimicrobial titanium/silver hard coating via the physical vapor deposition (PVD) process.

    METHODS:

    Coatings with a thickness of approximately 2 mum were deposited on titanium surfaces by simultaneous vaporisation of both metals in an inert argon atmosphere with a silver content of approximately 0.7-9% as indicated by energy dispersive X-ray analysis. On these surfaces microorganisms and eukaryotic culture cells were grown.

    RESULTS:

    The coatings released sufficient silver ions (0.5-2.3 ppb) when immersed in PBS and showed significant antimicrobial potency against Staphylococcus epidermis and Klebsiella pneumoniae strains. At the same time, no cytotoxic effects of the coatings on osteoblast and epithelial cells were found.

    CONCLUSION:

    Due to similar mechanical performance when compared to pure titanium, the TiAg coatings should be suitable to provide antimicrobial activity on load-bearing implant surfaces.

    PMID:
    16556327
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC1435898
    Free PMC Article

    Images from this publication.See all images (6) Free text