Display Settings:

Format

Send to:

Choose Destination
See comment in PubMed Commons below
Opt Lett. 2006 Feb 15;31(4):480-2.

Fourier-transform coherent anti-Stokes Raman scattering microscopy.

Author information

  • 1Laboratoire d'Optique et Biosciences, Centre National de la Recherche Scientifique Unité Mixte de Recherche 7645, Institut National de la Santé et de la Recherche Médicale U696, Ecole Polytechnique, 91128 Palaiseau, France. jogilvie@umich.edu

Abstract

We report a novel Fourier-transform-based implementation of coherent anti-Stokes Raman scattering (CARS) microscopy. The method employs a single femtosecond laser source and a Michelson interferometer to create two pulse replicas that are fed into a scanning multiphoton microscope. By varying the time delay between the pulses, we time-resolve the CARS signal, permitting easy removal of the nonresonant background while providing high resolution, spectrally resolved images of CARS modes over the laser bandwidth (approximately 1500 cm(-1)). We demonstrate the method by imaging polystyrene beads in solvent.