Liquid chromatography-electrospray ionization mass spectrometry for direct identification and quantification of iophenoxic acid in serum

J Chromatogr B Analyt Technol Biomed Life Sci. 2006 Feb 17;832(1):144-57. doi: 10.1016/j.jchromb.2005.12.017. Epub 2006 Jan 18.

Abstract

A liquid chromatographic-electrospray ionization mass spectrometric technique was developed for direct quantitation of iophenoxic acid (IA) in serum. IA was spiked into canine, feline, bovine, equine, and porcine sera, extracted, and quantified using negative ion monitoring following chromatographic separation on a Luna C18(2) 3 microm (100 mm x 2.1mm) reversed-phase column. The limit of detection was 25 ng/mL and the limit of quantification was 50 ng/mL. Inter- and intra-assay accuracy (86-113% and 87-115%, respectively) and precision (1.8-7.7%) were calculated. Analysis of serum collected from feral pigs, raccoons, and opossums following ingestion of IA-marked baits confirmed the appropriateness of this method for bait acceptance studies.

MeSH terms

  • Animals
  • Chromatography, Liquid / methods*
  • Iopanoic Acid / analogs & derivatives*
  • Iopanoic Acid / analysis
  • Opossums
  • Raccoons
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Swine

Substances

  • iophenoxic acid
  • Iopanoic Acid