The HUKM Spinal Instrumentation System for adolescent idiopathic scoliosis. A biomechanical comparison study using finite element analysis

Med J Malaysia. 2005 Jul:60 Suppl C:30-4.

Abstract

This finite element analysis is aimed at comparing relative stiffness of three different posterior instrumentation constructs: the Hospital Universiti Kebangsaan Malaysia Spinal Instrumentation System (HUKM-SIS), the Cotrell-Dubousset Instrumentation (CDI) and Harrington Instrumentation System (HIS), used in the treatment of adolescent idiopathic scoliosis (AIS). The constructs were tested under various loads using MSC Patran 2001 r2a. Under increasing flexion loads, there was a linearly corresponding increase in deflection magnitudes for all constructs on the load-deflection curve. The CDI was the stiffest construct under axial, forward flexion and extension loads, followed by the HUKM-SIS and HIS. Under lateral bending loads, the HUKM-SIS construct was the stiffest followed by CDI and HIS. The HUKM-SIS construct was stiffer than HIS under torsional loads. We conclude that multiple pedicle screws increase the stiffness of posterior instrumentation constructs under all loads and inter-segmental spinous processes wiring increase the stiffness against lateral bending.

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Biomechanical Phenomena
  • Finite Element Analysis
  • Humans
  • Internal Fixators*
  • Materials Testing*
  • Scoliosis / surgery*