    Pediatr Crit Care Med. 2005 Nov;6(6):671-5.

    Evaluation of tissue saturation as a noninvasive measure of mixed venous saturation in children.

    Source

    Department of Anesthesiology and Critical Care Medicine and Division of Cardiology, The Children's Hospital of Philadelphia, Philadelphia, PA, USA.

    Abstract

    BACKGROUND:

    Mixed venous saturation (S & OV0456;o2) is an important measurement that helps guide the care of critically ill patients. Invasive S & OV0456;o2 assessment in infants and children is often avoided because of the inherent risks. A noninvasive tissue saturation (S to 2) monitor has recently been developed that uses near-infrared spectroscopy to measure oxyhemoglobin saturation in muscle. In adult and animal studies, S to 2 correlated with oxygen delivery and S & OV0456;o2. There has been no evaluation in pediatric patients.

    OBJECTIVE:

    To evaluate tissue saturation as a noninvasive measure of mixed venous saturation in children.

    DESIGN:

    A prospective observational study.

    SETTING:

    Catheterization laboratory in a tertiary care children's medical center.

    PATIENTS:

    We studied 98 children (49 without intracardiac mixing and 49 with intracardiac mixing) <or=12 yrs of age who underwent cardiac catheterization. Under general anesthesia, we compared S to 2 measured over the deltoid muscle with superior vena cava saturation in both groups and S to 2 with pulmonary artery saturation in patients without intracardiac mixing. Paired measurements were analyzed for bias, precision, and correlation via Bland-Altman plot and linear regression.

    RESULTS:

    No meaningful correlation was found between S to 2 and superior vena cava saturation or pulmonary artery saturation. Bland-Altman analyses of S to 2 with superior vena cava saturation yielded bias values of -6.67 +/- 37.33% in patients with intracardiac mixing and -0.82 +/- 41.31% in patients without mixing. Bland-Altman analysis of S to 2 with pulmonary artery saturation yielded a bias of 3.61 +/- 41.32% in patients without mixing. Differences between noninvasive and invasive measurements were greatest in smaller children.

    CONCLUSION:

    Noninvasive tissue saturation over the deltoid does not correlate with S & OV0456;o2 in children. It is possible that more precise probe spacing, coupled with optimal muscle-mass location, could result in more accurate measures in future investigations.

    PMID:
    16276334
    [PubMed - indexed for MEDLINE]