    Ultrasound Med Biol. 2005 Oct;31(10):1343-9.

    In vivo volumetric analysis of coronary stent using optical coherence tomography with a novel balloon occlusion-flushing catheter: a comparison with intravascular ultrasound.

    Source

    Cardiology Laboratory for Integrative Physiology and Imaging, Massachusetts General Hospital, Harvard Medical School, Boston, MA, USA. ykawase@partners.org

    Abstract

    Optical coherence tomography (OCT) is limited as an intravascular imaging tool because of interference with blood. This study tested a new balloon occlusion-flushing catheter for OCT scanning of stented coronary arteries and compared stent measurements between OCT and intravascular ultrasound (IVUS). Motorized pullback with OCT and IVUS was examined in coronary stents deployed in swine. Quantitative measurements were obtained and compared between both groups. In addition, stent strut thickness was compared among OCT, IVUS and actual measurement. The occlusion catheter successfully provided motorized pullback OCT images in the stented coronary arteries without any complications. There were no differences in calculated lumen volume. However, stent volumes were significantly smaller with OCT than with IVUS (p < 0.05). OCT significantly underestimated the stent strut thickness compared with the actual measurement. Although OCT underestimates the stent strut thickness, motorized pullback OCT imaging with the occlusion catheter can provide appropriate in-stent images in the porcine coronary arteries.

    PMID:
    16223637
    [PubMed - indexed for MEDLINE]

      Supplemental Content

      Icon for Elsevier Science

      Save items

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk