    J Clin Pathol. 2005 Oct;58(10):1086-90.

    IHC for Her2 with CBE356 antibody is a more accurate predictor of Her2 gene amplification by FISH than HercepTest in breast carcinoma.

    Source

    Department of Pathology, Glasgow Royal Infirmary, Glasgow G4 0SF, UK.

    Abstract

    BACKGROUND:

    Her2 (c-erbB-2/neu) overexpression in breast carcinoma predicts response to the anti-Her2 monoclonal antibody, trastuzumab, and is associated with a poor prognosis. When considering patients for trastuzumab treatment, Her2 protein expression is measured by imunohistochemistry (IHC) and, where staining is equivocal, by fluorescence in situ hybridisation (FISH) detection of Her2 gene amplification.

    AIMS:

    To compare IHC using CBE356 with IHC using the Food and Drug Administration approved HercepTesttrade mark.

    METHODS:

    CBE356 and HercepTest were analysed using 167 FISH characterised breast carcinomas. Immunohistochemical expression of Her2 was measured semiquantitatively. Sensitivity, specificity, predictive values, and overall accuracy were calculated for both IHC methods using gene amplification by FISH as the end point, and IHC and FISH assays were tested in Kaplan-Meier survival analysis.

    RESULTS:

    The accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of CBE356 positive (2+ and 3+) cases were 94%, 89%, 95%, 84%, and 97%, respectively, and of HercepTest positive (2+ and 3+) cases were 91%, 66%, 98%, 92%, and 91%, respectively. A positive result with CBE356, HercepTest, or FISH was associated with significantly decreased overall survival (log rank p = 0.005, p = 0.0017, and p = 0.0005, respectively).

    CONCLUSIONS:

    Positive IHC staining for Her2 using CBE356 is 3% more accurate and 23% more sensitive at predicting Her2 gene amplification by FISH than positive staining with HercepTest. Negative IHC using CBE356 antibody is 6% more likely to represent a truly negative result than negative staining with HercepTest. Overall, CBE356 was a more accurate predictor of Her2 gene amplification by FISH than HercepTest.

    PMID:
    16189156
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC1770743
    Free PMC Article

    Images from this publication.See all images (5) Free text

    Figure 1
    Figure 3
    Figure 5
    Figure 2
    Figure 4

      Supplemental Content

      Icon for HighWire Press Icon for PubMed Central

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk