Display Settings:

Format

Send to:

Choose Destination
Opt Lett. 2004 Nov 15;29(22):2611-3.

Measuring 0.1-nm motion in 1 ms in an optical microscope with differential back-focal-plane detection.

Author information

  • 1JILA, National Institute of Standards and Technology, Boulder, Colorado 80309-0440, USA.

Abstract

Back-focal-plane detection of micrometer-sized beads offers subnanometer resolution for single-molecule, optical trapping experiments. However, laser beam-pointing instability and mechanical drift of the microscope limit the resolution of optical-trapping experiments. By combining two infrared lasers with improved differential beam-pointing stability (< or = 0.05 microrad), we simultaneously measure and subtract the motion of the microscope stage, leading to a resolution of <0.1 nm in 1 ms and stability of 0.5 nm over 60 s. Repeated steps of 0.4 nm at 1 Hz are resolved with a signal-to-noise ratio of 25.

PMID:
15552661
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Optical Society of America
    Loading ...
    Write to the Help Desk