Format

Send to:

Choose Destination
See comment in PubMed Commons below
Med Phys. 2004 Sep;31(9):2466-70.

Brachytherapy dosimetry parameters calculated for a new 103Pd source.

Author information

  • 1Department of Radiation Oncology, Tufts University School of Medicine, Tufts-New England Medical Center, Box #246, 750 Washington Street, Boston, Massachusetts 02111, USA.

Abstract

A new brachytherapy source having 103Pd adsorbed onto silver beads has been designed. The dose distributions of this source have been characterized using version 5 of the MCNP Monte Carlo radiation transport code available from Oak Ridge National Laboratory. These results are presented in terms of the updated AAPM Task Group No. 43 (TG-43U1) formalism, dosimetry parameters, and recommended calculation methodology.

PMID:
15487726
[PubMed - indexed for MEDLINE]