    J Arthroplasty. 2004 Sep;19(6):775-82.

    The stability of the cemented tibial component of total knee arthroplasty: posterior cruciate-retaining versus posterior-stabilized design.

    Source

    Istituti Ortopedici, Rizzoli, Bologna, Italy.

    Abstract

    Micromotion of the tibial component in 40 knee arthroplasties for gonarthrosis was studied using Roentgen stereophotogrammetric analysis. The stability of this component was assessed for 2 years' postoperatively. In all arthroplasties, an attempt was made to reconstruct the preoperative posterior slope. Posterior cruciate-retaining (CR) and posterior-stabilized (PS) components showed at 2 years a maximum total point motion of 0.6 +/- 0.4 mm and 0.7 +/- 0.5 mm, respectively. Whereas 92.5% of the implants were determined to be stable, 1 of the CR group and 2 of the PS group displayed migration between the first and the second year of at least 0.2 mm. A negative correlation between subsidence of the tibial component at 2 years of follow-up and the difference between preoperative and postoperative tibial slope was found. Consequently, we suggest that restoring the original posterior slope of the tibial plateau must be a goal of tibial component implantation.