Geometric leaf placement strategies

Phys Med Biol. 2004 Apr 21;49(8):1505-19. doi: 10.1088/0031-9155/49/8/009.

Abstract

Geometric leaf placement strategies for multileaf collimators (MLCs) typically involve the expansion of the beam's-eye-view contour of a target by a uniform MLC margin, followed by movement of the leaves until some point on each leaf end touches the expanded contour. Film-based dose-distribution measurements have been made to determine appropriate MLC margins--characterized through an index d90--for multileaves set using one particular strategy to straight lines lying at various angles to the direction of leaf travel. Simple trigonometric relationships exist between different geometric leaf placement strategies and are used to generalize the results of the film work into d90 values for several different strategies. Measured d90 values vary both with angle and leaf placement strategy. A model has been derived that explains and describes quite well the observed variations of d90 with angle. The d90 angular variations of the strategies studied differ substantially, and geometric and dosimetric reasoning suggests that the best strategy is the one with the least angular variation. Using this criterion, the best straightforwardly implementable strategy studied is a 'touch circle' approach for which semicircles are imagined to be inscribed within leaf ends, the leaves being moved until the semicircles just touch the expanded target outline.

MeSH terms

  • Film Dosimetry / methods
  • Humans
  • Models, Theoretical
  • Particle Accelerators
  • Phantoms, Imaging
  • Radiometry / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Conformal / instrumentation*
  • Radiotherapy, Conformal / methods*
  • Radiotherapy, High-Energy / methods