1: J Chromatogr A. 2004 Feb 20;1027(1-2):75-82.Links

High-speed analysis of residual solvents by flow-modulation gas chromatography.

Restek Corporation, 110 Benner Circle, Bellefonte, PA 16823, USA.

High-speed gas chromatographic (GC) separation of residual solvents in pharmaceutical preparations, using a flow-modulation technique, is described. These volatile compounds are separated on a series-coupled (tandem) column ensemble consisting of a polyethylene glycol column and a trifluoropropylmethyl/dimethylpolysiloxane column. This column ensemble is operated in stop-flow mode to enhance, or "tune", the separation. A valve between the junction point of the tandem column ensemble and a source of carrier gas at a pressure above the GC inlet pressure is opened for intervals of 2-8 s. This stops or slightly reverses the flow of carrier gas in the first column. Stop-flow pulses are used to increase the separation of target analytes that overlap in the total ensemble chromatogram, compared to non-stop-flow, or conventional, operation. All 36 target compounds, based on ICH Classes I and II residual solvent lists, are resolved in 12 min using the stop-flow technique and a single chromatographic analysis.

PMID: 14971486 [PubMed - indexed for MEDLINE]