    Nucleic Acids Res. 2004 Jan 23;32(2):e18.

    Rapid detection of single nucleotide polymorphisms associated with spinal muscular atrophy by use of a reusable fibre-optic biosensor.

    Source

    Chemical Sensors Group, University of Toronto at Mississauga, 3359 Mississauga Road North, Mississauga, Ontario, Canada.

    Abstract

    Rapid (<2 min) and quantitative genotyping for single nucleotide polymorphisms (SNPs) associated with spinal muscular atrophy was done using a reusable (approximately 80 cycles of application) fibre-optic biosensor over a clinically relevant range (0-4 gene copies). Sensors were functionalized with oligonucleotide probes immobilized at high density (approximately 7 pmol/cm2) to impart enhanced selectivity for SNP discrimination and used in a total internal reflection fluorescence detection motif to detect 202 bp PCR amplicons from patient samples. Real-time detection may be done over a range of ionic strength conditions (0.1-1.0 M) without stringency rinsing to remove non-selectively bound materials and without loss of selectivity, permitting a means for facile sample preparation. By using the time-derivative of fluorescence intensity as the analytical parameter, linearity of response may be maintained while allowing for significant reductions in analysis time (10-100-fold), permitting for the completion of measurements in under 1 min.

    PMID:
    14742865
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC373370
    Free PMC Article

    Images from this publication.See all images (6) Free text

    Figure 2
    Figure 4
    Figure 6
    Figure 1
    Figure 3
    Figure 5

      Supplemental Content

      Icon for HighWire Press Icon for PubMed Central

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk