    Rofo. 1977 Jun;126(6):529-36.

    [Image quality in film and xero-mammography. I. Rendering of low contrast values (author's transl)].

    [Article in German]

    Abstract

    The rendering of low contrast values by various mammographic imaging systems was investigated by a semi-objective procedure. Test exposures of a line raster with variable line widths were evaluated visually. Significantly better results were achieved by a xero-radiographic method. This was better than non-screen mammographic film in terms of object size, lower sensitivity to scatter and better resolution. The use of a tungsten rotating anode, with about 0.5 mm. Al filter, in xeromammography provided a satisfactory compromise between image quality and radiation dose in the rendering of low contrast studies.